CVE-2025-54487
CRITICALDescription
A stack-based buffer overflow vulnerability exists in the MFER parsing functionality of The Biosig Project libbiosig 3.9.0 and Master Branch (35a819fa). A specially crafted MFER file can lead to arbitrary code execution. An attacker can provide a malicious file to trigger this vulnerability.This vulnerability manifests on line 8842 of biosig.c on the current master branch (35a819fa), when the Tag is 12: else if (tag==12) //0x0C { // sampling resolution if (len>6) fprintf(stderr,"Warning MFER tag12 incorrect length %i>6\n",len); val32 = 0; int8_t v8; curPos += ifread(&UnitCode,1,1,hdr); curPos += ifread(&v8,1,1,hdr); curPos += ifread(buf,1,len-2,hdr); In addition to values of `len` greater than 130 triggering a buffer overflow, a value of `len` smaller than 2 will also trigger a buffer overflow due to an integer underflow when computing `len-2` in this code path.
CVE Details
Affected Products
Weaknesses (CWE)
References
IOC Correlations
No correlations recorded
This product uses data from the NVD API but is not endorsed or certified by the NVD.