TROYANOSYVIRUS
Back to CVEs

CVE-2023-25931

MEDIUM
6.4

Description

Medtronic identified that the Pelvic Health clinician apps, which are installed on the Smart Programmer mobile device, have a password vulnerability that requires a security update to fix. Not updating could potentially result in unauthorized control of the clinician therapy application, which has greater control over therapy parameters than the patient app. Changes still cannot be made outside of the established therapy parameters of the programmer. For unauthorized access to occur, an individual would need physical access to the Smart Programmer.

CVE Details

CVSS v3.1 Score6.4
SeverityMEDIUM
CVSS VectorCVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:L/I:H/A:H
Attack VectorPHYSICAL
ComplexityLOW
Privileges RequiredNONE
User InteractionNONE
Published3/1/2023
Last Modified11/21/2024
Sourcenvd
Honeypot Sightings0

Affected Products

medtronic:interstim_x_clinicianmedtronic:micro_clinician

Weaknesses (CWE)

CWE-620CWE-287

IOC Correlations

No correlations recorded

This product uses data from the NVD API but is not endorsed or certified by the NVD.