Description
Sigma Spectrum Infusion System v's6.x (model 35700BAX) and Baxter Spectrum Infusion System Version(s) 8.x (model 35700BAX2) at the application layer uses an unauthenticated clear-text communication channel to send and receive system status and operational data. This could allow an attacker that has circumvented network security measures to view sensitive non-private data or to perform a man-in-the-middle attack.
CVE Details
CVSS v3.1 Score9.8
SeverityCRITICAL
CVSS VectorCVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
Attack VectorNETWORK
ComplexityLOW
Privileges RequiredNONE
User InteractionNONE
Published6/29/2020
Last Modified11/21/2024
Sourcenvd
Honeypot Sightings0
Affected Products
baxter:sigma_spectrum_infusion_systembaxter:sigma_spectrum_infusion_system_firmware
Weaknesses (CWE)
CWE-319CWE-319
References
https://www.us-cert.gov/ics/advisories/icsma-20-170-04(ics-cert@hq.dhs.gov)
https://www.us-cert.gov/ics/advisories/icsma-20-170-04(af854a3a-2127-422b-91ae-364da2661108)
IOC Correlations
No correlations recorded