← Back to CVEs
CVE-2018-10631
MEDIUM6.3
Description
The 8840 Clinician Programmer executes the application program from the 8870 Application Card. An attacker with physical access to an 8870 Application Card and sufficient technical capability can modify the contents of this card, including the binary executables. If modified to bypass protection mechanisms, this malicious code will be run when the card is inserted into an 8840 Clinician Programmer.
CVE Details
CVSS v3.1 Score6.3
SeverityMEDIUM
CVSS VectorCVSS:3.1/AV:P/AC:H/PR:N/UI:R/S:U/C:H/I:H/A:H
Attack VectorPHYSICAL
ComplexityHIGH
Privileges RequiredNONE
User InteractionREQUIRED
Published7/13/2018
Last Modified8/26/2025
Sourcenvd
Honeypot Sightings0
Affected Products
medtronic:n\'vision_8840medtronic:n\'vision_8840_firmwaremedtronic:n\'vision_8870medtronic:n\'vision_8870_firmware
Weaknesses (CWE)
CWE-693CWE-693
References
http://www.securityfocus.com/bid/104213(ics-cert@hq.dhs.gov)
https://global.medtronic.com/xg-en/product-security/security-bulletins/nvision.html(ics-cert@hq.dhs.gov)
https://ics-cert.us-cert.gov/advisories/ICSMA-18-137-01(ics-cert@hq.dhs.gov)
https://www.medtronic.com/security(ics-cert@hq.dhs.gov)
https://ics-cert.us-cert.gov/advisories/ICSMA-18-137-01(af854a3a-2127-422b-91ae-364da2661108)
https://www.medtronic.com/security(af854a3a-2127-422b-91ae-364da2661108)
IOC Correlations
No correlations recorded
This product uses data from the NVD API but is not endorsed or certified by the NVD.