N/ADescription
The Hospira LifeCare PCA Infusion System before 7.0 does not validate network traffic associated with sending a (1) drug library, (2) software update, or (3) configuration change, which allows remote attackers to modify settings or medication data via packets on the (a) TELNET, (b) HTTP, (c) HTTPS, or (d) UPNP port. NOTE: this issue might overlap CVE-2015-3459.
CVE Details
CVSS v3.1 ScoreN/A
Published7/6/2015
Last Modified11/3/2025
Sourcenvd
Honeypot Sightings0
Affected Products
hospira:lifecare_pca3hospira:lifecare_pca5hospira:lifecare_pcainfusion_firmware
Weaknesses (CWE)
CWE-345CWE-345
References
http://www.fda.gov/MedicalDevices/Safety/AlertsandNotices/ucm446809.htm(ics-cert@hq.dhs.gov)
https://github.com/cisagov/CSAF/blob/develop/csaf_files/OT/white/2015/icsa-15-125-01.json(ics-cert@hq.dhs.gov)
https://www.cisa.gov/news-events/ics-advisories/icsa-15-125-01(ics-cert@hq.dhs.gov)
https://xs-sniper.com/blog/2015/06/08/hospira-plum-a-infusion-pump-vulnerabilities/(ics-cert@hq.dhs.gov)
http://www.fda.gov/MedicalDevices/Safety/AlertsandNotices/ucm446809.htm(af854a3a-2127-422b-91ae-364da2661108)
https://ics-cert.us-cert.gov/advisories/ICSA-15-125-01(af854a3a-2127-422b-91ae-364da2661108)
https://xs-sniper.com/blog/2015/06/08/hospira-plum-a-infusion-pump-vulnerabilities/(af854a3a-2127-422b-91ae-364da2661108)