CVE-2024-48966
CRITICAL10.0
Descricao
The software tools used by service personnel to test & calibrate the ventilator do not support user authentication. An attacker with access to the Service PC where the tools are installed could obtain diagnostic information through the test tool or manipulate the ventilator's settings and embedded software via the calibration tool, without having to authenticate to either tool. This could result in unauthorized disclosure of information and/or have unintended impacts on device settings and performance.
Detalhes CVE
Pontuacao CVSS v3.110.0
SeveridadeCRITICAL
Vetor CVSSCVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H
Vetor de ataqueNETWORK
ComplexidadeLOW
Privilegios necessariosNONE
Interacao do usuarioNONE
Publicado11/14/2024
Ultima modificacao11/15/2024
Fontenvd
Avistamentos honeypot0
Fraquezas (CWE)
CWE-306
Referencias
https://www.cisa.gov/news-events/ics-medical-advisories/icsma-24-319-01(productsecurity@baxter.com)