TROYANOSYVIRUS
Voltar para CVEs

CVE-2023-25931

MEDIUM
6.4

Descricao

Medtronic identified that the Pelvic Health clinician apps, which are installed on the Smart Programmer mobile device, have a password vulnerability that requires a security update to fix. Not updating could potentially result in unauthorized control of the clinician therapy application, which has greater control over therapy parameters than the patient app. Changes still cannot be made outside of the established therapy parameters of the programmer. For unauthorized access to occur, an individual would need physical access to the Smart Programmer.

Detalhes CVE

Pontuacao CVSS v3.16.4
SeveridadeMEDIUM
Vetor CVSSCVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:L/I:H/A:H
Vetor de ataquePHYSICAL
ComplexidadeLOW
Privilegios necessariosNONE
Interacao do usuarioNONE
Publicado3/1/2023
Ultima modificacao11/21/2024
Fontenvd
Avistamentos honeypot0

Produtos afetados

medtronic:interstim_x_clinicianmedtronic:micro_clinician

Fraquezas (CWE)

CWE-620CWE-287

Correlacoes IOC

Sem correlacoes registradas

This product uses data from the NVD API but is not endorsed or certified by the NVD.