TROYANOSYVIRUS
Voltar para CVEs

CVE-2023-0888

MEDIUM
4.9

Descricao

An improper neutralization of directives in dynamically evaluated code vulnerability in the WiFi Battery embedded web server in versions L90/U70 and L92/U92 can be used to gain administrative access to the WiFi communication module. An authenticated user, having access to both the medical device WiFi network (such as a biomedical engineering staff member) and the specific B.Braun Battery Pack SP with WiFi web server credentials, could get administrative (root) access on the infusion pump communication module. This could be used as a vector to start further attacks

Detalhes CVE

Pontuacao CVSS v3.14.9
SeveridadeMEDIUM
Vetor CVSSCVSS:3.1/AV:N/AC:L/PR:H/UI:N/S:U/C:N/I:H/A:N
Vetor de ataqueNETWORK
ComplexidadeLOW
Privilegios necessariosHIGH
Interacao do usuarioNONE
Publicado3/13/2023
Ultima modificacao11/21/2024
Fontenvd
Avistamentos honeypot0

Produtos afetados

bbraun:battery-pack_sp_with_wifibbraun:battery-pack_sp_with_wifi_firmware

Fraquezas (CWE)

CWE-95CWE-94

Referencias

https://www.bbraun.com/productsecurity(productsecurity@bbraun.com)
https://www.bbraunusa.com/productsecurity(productsecurity@bbraun.com)
https://www.bbraun.com/productsecurity(af854a3a-2127-422b-91ae-364da2661108)
https://www.bbraunusa.com/productsecurity(af854a3a-2127-422b-91ae-364da2661108)

Correlacoes IOC

Sem correlacoes registradas

This product uses data from the NVD API but is not endorsed or certified by the NVD.