← Voltar para CVEs
CVE-2021-37162
CRITICAL9.8
Descricao
A buffer overflow issue was discovered in HMI3 Control Panel in Swisslog Healthcare Nexus Panel operated by released versions of software before Nexus Software 7.2.5.7. If an attacker sends a malformed UDP message, a buffer underflow occurs, leading to an out-of-bounds copy and possible remote code execution.
Detalhes CVE
Pontuacao CVSS v3.19.8
SeveridadeCRITICAL
Vetor CVSSCVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
Vetor de ataqueNETWORK
ComplexidadeLOW
Privilegios necessariosNONE
Interacao do usuarioNONE
Publicado8/2/2021
Ultima modificacao11/21/2024
Fontenvd
Avistamentos honeypot0
Produtos afetados
swisslog-healthcare:hmi-3_control_panelswisslog-healthcare:hmi-3_control_panel_firmware
Fraquezas (CWE)
CWE-120
Referencias
https://www.armis.com/PwnedPiper(cve@mitre.org)
https://www.swisslog-healthcare.com(cve@mitre.org)
https://www.armis.com/PwnedPiper(af854a3a-2127-422b-91ae-364da2661108)
https://www.swisslog-healthcare.com(af854a3a-2127-422b-91ae-364da2661108)
Correlacoes IOC
Sem correlacoes registradas
This product uses data from the NVD API but is not endorsed or certified by the NVD.