← Voltar para CVEs
CVE-2018-10631
MEDIUM6.3
Descricao
The 8840 Clinician Programmer executes the application program from the 8870 Application Card. An attacker with physical access to an 8870 Application Card and sufficient technical capability can modify the contents of this card, including the binary executables. If modified to bypass protection mechanisms, this malicious code will be run when the card is inserted into an 8840 Clinician Programmer.
Detalhes CVE
Pontuacao CVSS v3.16.3
SeveridadeMEDIUM
Vetor CVSSCVSS:3.1/AV:P/AC:H/PR:N/UI:R/S:U/C:H/I:H/A:H
Vetor de ataquePHYSICAL
ComplexidadeHIGH
Privilegios necessariosNONE
Interacao do usuarioREQUIRED
Publicado7/13/2018
Ultima modificacao8/26/2025
Fontenvd
Avistamentos honeypot0
Produtos afetados
medtronic:n\'vision_8840medtronic:n\'vision_8840_firmwaremedtronic:n\'vision_8870medtronic:n\'vision_8870_firmware
Fraquezas (CWE)
CWE-693CWE-693
Referencias
http://www.securityfocus.com/bid/104213(ics-cert@hq.dhs.gov)
https://global.medtronic.com/xg-en/product-security/security-bulletins/nvision.html(ics-cert@hq.dhs.gov)
https://ics-cert.us-cert.gov/advisories/ICSMA-18-137-01(ics-cert@hq.dhs.gov)
https://www.medtronic.com/security(ics-cert@hq.dhs.gov)
https://ics-cert.us-cert.gov/advisories/ICSMA-18-137-01(af854a3a-2127-422b-91ae-364da2661108)
https://www.medtronic.com/security(af854a3a-2127-422b-91ae-364da2661108)
Correlacoes IOC
Sem correlacoes registradas
This product uses data from the NVD API but is not endorsed or certified by the NVD.