TROYANOSYVIRUS
Retour aux CVEs

CVE-2023-0888

MEDIUM
4.9

Description

An improper neutralization of directives in dynamically evaluated code vulnerability in the WiFi Battery embedded web server in versions L90/U70 and L92/U92 can be used to gain administrative access to the WiFi communication module. An authenticated user, having access to both the medical device WiFi network (such as a biomedical engineering staff member) and the specific B.Braun Battery Pack SP with WiFi web server credentials, could get administrative (root) access on the infusion pump communication module. This could be used as a vector to start further attacks

Details CVE

Score CVSS v3.14.9
SeveriteMEDIUM
Vecteur CVSSCVSS:3.1/AV:N/AC:L/PR:H/UI:N/S:U/C:N/I:H/A:N
Vecteur d'attaqueNETWORK
ComplexiteLOW
Privileges requisHIGH
Interaction utilisateurNONE
Publie3/13/2023
Derniere modification11/21/2024
Sourcenvd
Observations honeypot0

Produits affectes

bbraun:battery-pack_sp_with_wifibbraun:battery-pack_sp_with_wifi_firmware

Faiblesses (CWE)

CWE-95CWE-94

References

https://www.bbraun.com/productsecurity(productsecurity@bbraun.com)
https://www.bbraunusa.com/productsecurity(productsecurity@bbraun.com)
https://www.bbraun.com/productsecurity(af854a3a-2127-422b-91ae-364da2661108)
https://www.bbraunusa.com/productsecurity(af854a3a-2127-422b-91ae-364da2661108)

Correlations IOC

Aucune correlation enregistree