← Retour aux CVEs
CVE-2021-38392
MEDIUM6.5
Description
A skilled attacker with physical access to the affected device can gain access to the hard disk drive of the device to change the telemetry region and could use this setting to interrogate or program an implantable device in any region in the world.
Details CVE
Score CVSS v3.16.5
SeveriteMEDIUM
Vecteur CVSSCVSS:3.1/AV:P/AC:L/PR:N/UI:R/S:C/C:L/I:H/A:L
Vecteur d'attaquePHYSICAL
ComplexiteLOW
Privileges requisNONE
Interaction utilisateurREQUIRED
Publie10/4/2021
Derniere modification11/21/2024
Sourcenvd
Observations honeypot0
Produits affectes
bostonscientific:zoom_latitude_pogrammer\/recorder\/monitor_3120bostonscientific:zoom_latitude_pogrammer\/recorder\/monitor_3120_firmware
Faiblesses (CWE)
CWE-284
References
https://us-cert.cisa.gov/ics/advisories/icsma-21-273-01(ics-cert@hq.dhs.gov)
https://us-cert.cisa.gov/ics/advisories/icsma-21-273-01(af854a3a-2127-422b-91ae-364da2661108)
Correlations IOC
Aucune correlation enregistree