Description

An improper sanitization of input vulnerability in B. Braun SpaceCom2 prior to 012U000062 allows a remote unauthenticated attacker to gain user-level command-line access by passing a raw external string straight through to printf statements. The attacker is required to be on the same network as the device.

Details CVE

Score CVSS v3.18.1
SeveriteHIGH
Vecteur CVSSCVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:N
Vecteur d'attaqueADJACENT_NETWORK
ComplexiteLOW
Privileges requisNONE
Interaction utilisateurNONE
Publie8/25/2021
Derniere modification11/21/2024
Sourcenvd
Observations honeypot0

Produits affectes

bbraun:infusomat_large_volume_pump_871305ubbraun:spacecom2bbraun:spacestation_8713142u

Faiblesses (CWE)

CWE-134