CVE-2020-25156
HIGH7.2
Description
Active debug code in the B. Braun Melsungen AG SpaceCom Version L8/U61, and the Data module compactplus Versions A10 and A11 and earlier enables attackers in possession of cryptographic material to access the device as root.
Details CVE
Score CVSS v3.17.2
SeveriteHIGH
Vecteur CVSSCVSS:3.1/AV:N/AC:L/PR:H/UI:N/S:U/C:H/I:H/A:H
Vecteur d'attaqueNETWORK
ComplexiteLOW
Privileges requisHIGH
Interaction utilisateurNONE
Publie4/14/2022
Derniere modification11/21/2024
Sourcenvd
Observations honeypot0
Produits affectes
bbraun:datamodule_compactplusbbraun:spacecom
Faiblesses (CWE)
CWE-489
References
https://www.bbraun.com/en/products-and-therapies/services/b-braun-vulnerability-disclosure-policy/security-advisory.html(ics-cert@hq.dhs.gov)
https://www.cisa.gov/uscert/ics/advisories/icsma-20-296-02(ics-cert@hq.dhs.gov)
https://www.bbraun.com/en/products-and-therapies/services/b-braun-vulnerability-disclosure-policy/security-advisory.html(af854a3a-2127-422b-91ae-364da2661108)
https://www.cisa.gov/uscert/ics/advisories/icsma-20-296-02(af854a3a-2127-422b-91ae-364da2661108)