Description

Baxter ExactaMix EM 2400 & EM 1200, Versions ExactaMix EM2400 Versions 1.10, 1.11, 1.13, 1.14, ExactaMix EM1200 Versions 1.1, 1.2, 1.4, 1.5, Baxter ExactaMix EM 2400 Versions 1.10, 1.11, 1.13, 1.14 and ExactaMix EM1200 Versions 1.1, 1.2, 1.4 and 1.5 have hard-coded administrative account credentials for the ExactaMix operating system. Successful exploitation of this vulnerability may allow an attacker who has gained unauthorized access to system resources, including access to execute software or to view/update files, directories, or system configuration. This could allow an attacker with network access to view sensitive data including PHI.

Details CVE

Score CVSS v3.19.8
SeveriteCRITICAL
Vecteur CVSSCVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
Vecteur d'attaqueNETWORK
ComplexiteLOW
Privileges requisNONE
Interaction utilisateurNONE
Publie6/29/2020
Derniere modification11/21/2024
Sourcenvd
Observations honeypot0

Produits affectes

baxter:em1200baxter:em1200_firmwarebaxter:em2400baxter:em2400_firmware

Faiblesses (CWE)

CWE-259CWE-798