CVE-2018-10631
MEDIUM6.3
Description
The 8840 Clinician Programmer executes the application program from the 8870 Application Card. An attacker with physical access to an 8870 Application Card and sufficient technical capability can modify the contents of this card, including the binary executables. If modified to bypass protection mechanisms, this malicious code will be run when the card is inserted into an 8840 Clinician Programmer.
Details CVE
Score CVSS v3.16.3
SeveriteMEDIUM
Vecteur CVSSCVSS:3.1/AV:P/AC:H/PR:N/UI:R/S:U/C:H/I:H/A:H
Vecteur d'attaquePHYSICAL
ComplexiteHIGH
Privileges requisNONE
Interaction utilisateurREQUIRED
Publie7/13/2018
Derniere modification8/26/2025
Sourcenvd
Observations honeypot0
Produits affectes
medtronic:n\'vision_8840medtronic:n\'vision_8840_firmwaremedtronic:n\'vision_8870medtronic:n\'vision_8870_firmware
Faiblesses (CWE)
CWE-693CWE-693
References
http://www.securityfocus.com/bid/104213(ics-cert@hq.dhs.gov)
https://global.medtronic.com/xg-en/product-security/security-bulletins/nvision.html(ics-cert@hq.dhs.gov)
https://ics-cert.us-cert.gov/advisories/ICSMA-18-137-01(ics-cert@hq.dhs.gov)
https://www.medtronic.com/security(ics-cert@hq.dhs.gov)
https://ics-cert.us-cert.gov/advisories/ICSMA-18-137-01(af854a3a-2127-422b-91ae-364da2661108)
https://www.medtronic.com/security(af854a3a-2127-422b-91ae-364da2661108)