← Zuruck zu CVEs
CVE-2024-48967
CRITICAL10.0
Beschreibung
The ventilator and the Service PC lack sufficient audit logging capabilities to allow for detection of malicious activity and subsequent forensic examination. An attacker with access to the ventilator and/or the Service PC could, without detection, make unauthorized changes to ventilator settings that result in unauthorized disclosure of information and/or have unintended impacts on device performance.
CVE Details
CVSS v3.1 Bewertung10.0
SchweregradCRITICAL
CVSS VektorCVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H
AngriffsvektorNETWORK
KomplexitatLOW
Erforderliche PrivilegienNONE
BenutzerinteraktionNONE
Veroffentlicht11/14/2024
Zuletzt geandert11/15/2024
Quellenvd
Honeypot-Sichtungen0
Schwachen (CWE)
CWE-778
Referenzen
https://www.cisa.gov/news-events/ics-medical-advisories/icsma-24-319-01(productsecurity@baxter.com)
IOC Korrelationen
Keine Korrelationen erfasst