← Zuruck zu CVEs
CVE-2021-33885
CRITICAL10.0
Beschreibung
An Insufficient Verification of Data Authenticity vulnerability in B. Braun SpaceCom2 prior to 012U000062 allows a remote unauthenticated attacker to send the device malicious data that will be used in place of the correct data. This results in full system command access and execution because of the lack of cryptographic signatures on critical data sets.
CVE Details
CVSS v3.1 Bewertung10.0
SchweregradCRITICAL
CVSS VektorCVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:N
AngriffsvektorNETWORK
KomplexitatLOW
Erforderliche PrivilegienNONE
BenutzerinteraktionNONE
Veroffentlicht8/25/2021
Zuletzt geandert11/21/2024
Quellenvd
Honeypot-Sichtungen0
Betroffene Produkte
bbraun:infusomat_large_volume_pump_871305ubbraun:spacecom2bbraun:spacestation_8713142u
Schwachen (CWE)
CWE-347
Referenzen
https://www.bbraunusa.com/en.htm(cve@mitre.org)
https://www.bbraunusa.com/en.htm(af854a3a-2127-422b-91ae-364da2661108)
https://www.mcafee.com/blogs/enterprise/mcafee-enterprise-atr/mcafee-enterprise-atr-uncovers-vulnerabilities-in-globally-used-b-braun-infusion-pump/(af854a3a-2127-422b-91ae-364da2661108)
IOC Korrelationen
Keine Korrelationen erfasst
This product uses data from the NVD API but is not endorsed or certified by the NVD.