Beschreibung

Sigma Spectrum Infusion System v's6.x (model 35700BAX) and Baxter Spectrum Infusion System Version(s) 8.x (model 35700BAX2) at the application layer uses an unauthenticated clear-text communication channel to send and receive system status and operational data. This could allow an attacker that has circumvented network security measures to view sensitive non-private data or to perform a man-in-the-middle attack.

CVE Details

CVSS v3.1 Bewertung9.8
SchweregradCRITICAL
CVSS VektorCVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
AngriffsvektorNETWORK
KomplexitatLOW
Erforderliche PrivilegienNONE
BenutzerinteraktionNONE
Veroffentlicht6/29/2020
Zuletzt geandert11/21/2024
Quellenvd
Honeypot-Sichtungen0

Betroffene Produkte

baxter:sigma_spectrum_infusion_systembaxter:sigma_spectrum_infusion_system_firmware

Schwachen (CWE)

CWE-319CWE-319