TROYANOSYVIRUS
Zuruck zu CVEs

CVE-2014-5432

N/A

Beschreibung

Baxter SIGMA Spectrum Infusion System version 6.05 (model 35700BAX) with wireless battery module (WBM) version 16 is remotely accessible via Port 22/SSH without authentication. A remote attacker may be able to make unauthorized configuration changes to the WBM, as well as issue commands to access account credentials and shared keys. Baxter asserts that this vulnerability only allows access to features and functionality on the WBM and that the SIGMA Spectrum infusion pump cannot be controlled from the WBM. Baxter has released a new version of the SIGMA Spectrum Infusion System, Version 8, which incorporates hardware and software changes.

CVE Details

CVSS v3.1 BewertungN/A
Veroffentlicht3/26/2019
Zuletzt geandert11/21/2024
Quellenvd
Honeypot-Sichtungen0

Betroffene Produkte

baxter:sigma_spectrum_infusion_systembaxter:sigma_spectrum_infusion_system_firmwarebaxter:wireless_battery_module

Schwachen (CWE)

CWE-592CWE-287

IOC Korrelationen

Keine Korrelationen erfasst

This product uses data from the NVD API but is not endorsed or certified by the NVD.