The ventilator and the Service PC lack sufficient audit logging capabilities to allow for detection of malicious activity and subsequent forensic examination. An attacker with access to the ventilator and/or the Service PC could, without detection, make unauthorized changes to ventilator settings that result in unauthorized disclosure of information and/or have unintended impacts on device performance.
Detalles CVE
Puntuacion CVSS v3.110.0
SeveridadCRITICAL
Vector CVSSCVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H
Vector de ataqueNETWORK
ComplejidadLOW
Privilegios requeridosNONE
Interaccion usuarioNONE
Publicado11/14/2024
Ultima modificacion11/15/2024
Fuentenvd
Avistamientos honeypot0
Debilidades (CWE)
CWE-778
Referencias
https://www.cisa.gov/news-events/ics-medical-advisories/icsma-24-319-01(productsecurity@baxter.com)