TROYANOSYVIRUS
Volver a CVEs

CVE-2024-48966

CRITICAL
10.0

Descripcion

The software tools used by service personnel to test & calibrate the ventilator do not support user authentication. An attacker with access to the Service PC where the tools are installed could obtain diagnostic information through the test tool or manipulate the ventilator's settings and embedded software via the calibration tool, without having to authenticate to either tool. This could result in unauthorized disclosure of information and/or have unintended impacts on device settings and performance.

Detalles CVE

Puntuacion CVSS v3.110.0
SeveridadCRITICAL
Vector CVSSCVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H
Vector de ataqueNETWORK
ComplejidadLOW
Privilegios requeridosNONE
Interaccion usuarioNONE
Publicado11/14/2024
Ultima modificacion11/15/2024
Fuentenvd
Avistamientos honeypot0

Debilidades (CWE)

CWE-306

Correlaciones IOC

Sin correlaciones registradas