TROYANOSYVIRUS
Volver a CVEs

CVE-2023-0888

MEDIUM
4.9

Descripcion

An improper neutralization of directives in dynamically evaluated code vulnerability in the WiFi Battery embedded web server in versions L90/U70 and L92/U92 can be used to gain administrative access to the WiFi communication module. An authenticated user, having access to both the medical device WiFi network (such as a biomedical engineering staff member) and the specific B.Braun Battery Pack SP with WiFi web server credentials, could get administrative (root) access on the infusion pump communication module. This could be used as a vector to start further attacks

Detalles CVE

Puntuacion CVSS v3.14.9
SeveridadMEDIUM
Vector CVSSCVSS:3.1/AV:N/AC:L/PR:H/UI:N/S:U/C:N/I:H/A:N
Vector de ataqueNETWORK
ComplejidadLOW
Privilegios requeridosHIGH
Interaccion usuarioNONE
Publicado3/13/2023
Ultima modificacion11/21/2024
Fuentenvd
Avistamientos honeypot0

Productos afectados

bbraun:battery-pack_sp_with_wifibbraun:battery-pack_sp_with_wifi_firmware

Debilidades (CWE)

CWE-95CWE-94

Referencias

https://www.bbraun.com/productsecurity(productsecurity@bbraun.com)
https://www.bbraunusa.com/productsecurity(productsecurity@bbraun.com)
https://www.bbraun.com/productsecurity(af854a3a-2127-422b-91ae-364da2661108)
https://www.bbraunusa.com/productsecurity(af854a3a-2127-422b-91ae-364da2661108)

Correlaciones IOC

Sin correlaciones registradas

This product uses data from the NVD API but is not endorsed or certified by the NVD.