CVE-2021-33886
HIGH8.1
Descripcion
An improper sanitization of input vulnerability in B. Braun SpaceCom2 prior to 012U000062 allows a remote unauthenticated attacker to gain user-level command-line access by passing a raw external string straight through to printf statements. The attacker is required to be on the same network as the device.
Detalles CVE
Puntuacion CVSS v3.18.1
SeveridadHIGH
Vector CVSSCVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:N
Vector de ataqueADJACENT_NETWORK
ComplejidadLOW
Privilegios requeridosNONE
Interaccion usuarioNONE
Publicado8/25/2021
Ultima modificacion11/21/2024
Fuentenvd
Avistamientos honeypot0
Productos afectados
bbraun:infusomat_large_volume_pump_871305ubbraun:spacecom2bbraun:spacestation_8713142u
Debilidades (CWE)
CWE-134