Descripcion

Sigma Spectrum Infusion System v's6.x (model 35700BAX) and Baxter Spectrum Infusion System Version(s) 8.x (model 35700BAX2) at the application layer uses an unauthenticated clear-text communication channel to send and receive system status and operational data. This could allow an attacker that has circumvented network security measures to view sensitive non-private data or to perform a man-in-the-middle attack.

Detalles CVE

Puntuacion CVSS v3.19.8
SeveridadCRITICAL
Vector CVSSCVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
Vector de ataqueNETWORK
ComplejidadLOW
Privilegios requeridosNONE
Interaccion usuarioNONE
Publicado6/29/2020
Ultima modificacion11/21/2024
Fuentenvd
Avistamientos honeypot0

Productos afectados

baxter:sigma_spectrum_infusion_systembaxter:sigma_spectrum_infusion_system_firmware

Debilidades (CWE)

CWE-319CWE-319

Correlaciones IOC

Sin correlaciones registradas